Reduce Scrapping and Smelting, Promote Component Reuse   >   Electric Powered   >   Food Processing   >   DuPage County   >   Atn agilent atn-4111A s-parameter ~ 8753D 8753E 85047A

Atn agilent atn-4111A s-parameter ~ 8753D 8753E 85047A


ATN AGILENT ATN-4111A S-PARAMETER ~ 8753D 8753E 85047A
The ATN-4111A Multiport S-Parameter Test System allows true multiport vector-error corrected s-parameter measurements of both active and passive circuits. This test system contributes two-port accuracy and ease of use to reduce the errors, time and cost of measuring transistors, multi-stage amplifiers, differential amplifiers, couplers, hybrids, circulators, switches, multiplexers, transmission lines, and other multiport components. The ATN-4111A eliminates the manual manipulation of the Round Robin technique and delivers accuracy similar to a two-port VNA calibration. This is accomplished with the four-port test set hardware and companion application software that uses a multiple-term calibration algorithm to achieve true four-port error correction. The ATN-4111A four-port test set replaces the network analyzer two-port test set; multiplexing the A, B, and R channels to the four test ports. The system can measure the complete s-parameters of one-, two-, three-, and four-port devices.
* 4-port measurement capability
* 2-port accuracy and ease of use
* Practical, fast, single connection measurement
* Convenient, thorough, accurate measurement of multiport and differential devices
This unit supports the following equipment:
* HP 8753C (for measurements up to 3 GHz, firmware rev. 04.13 or higher)
* Pentium (or better) PC compatible computer
* Windows 95 or Windows NT
* At least 64 MBs of RAM
* HP 85050B or HP 85050C (7 mm)
* Additional calibration kit information may be added by the user
* HP 82340 or HP 82341 only
This system has the following specifications:
* Transmission measurements at 10 Hz IF bandwidth with full error correction and +5 dBm maximum output power.
* 50 MHz to 4 GHz: 90 dB
* Measurement Port Characteristics
* Frequency range: 05-2 GHz; 2-4 GHz; 4-6 GHz
* Directivity: 30 dB; 24 dB; 28 dB
* Source Match: 18 dB; 16 dB; 16 dB
* Load Match: 22 dB; 17 dB; 15 dB
* Refl. Tracking (+/-): 1.1 dB; .8 dB; 0.9 dB
* Trans. Tracking (+/-): 1.0 dB; 1.5 dB; 0.8 dB
* Frequency range: 50 MHz to 6.0 GHz
* Insertion loss, any test port-to-A/B: 50 MHz to 300 MHz: 22.4 dB max; 50 MHz to 6 GHz: 27.5 dB max.
* Insertion loss, RF-to-any test port: 50 MHz to 300 MHz: 7.7 dB max; 50 MHz to 6 GHz: 13.2 dB max.
* Insertion loss, RF-to-R: 50 MHz to 300 MHz: 29.0 dB max; 50 MHz to 6 GHz: 32.5 dB max.
* Isolation, port-to-port & A-to-B: >= 85 dB
* Maximum operating level: +20 dBm
* RF attenuator (option 001): 0-70 dB in 10 dB steps
* Test port connectors: 3.5 mm (m)
* VNA connectors: 50 ohm Type-N(f)
* 001 = (1) Integrated 70 dB (x 10 dB) step attenuator and (4) Integrated 0.5 Amp bias tees
* ATN Microwave ATN-4111A 50 MHz - 6 GHz S-paramter Test Set
* Multiport Professional Edition Software Ver. 1.16
This unit cannot be used without the software.